Analysis of 7 nitrobenzene in water samples by SPME-GC-TSDorECD

Your current location: Home page >> Case >> SPME application instance

Analysis of 7 nitrobenzene in water samples by SPME-GC-TSDorECD


  • Analysis of 7 nitrobenzene in water samples by SPME-GC-TSDorECD

Analysis of 7 nitrobenzene in water samples by SPME-GC-TSDorECD

Qingdao Zhenzheng Analytical Instrument Co., LTD. Technical department


SPMEProbe item numberZZ-SPME-06-s

 Analysis condition:                            图片1.png

 



  

Samples: 10 ml water samples with 2 ng/ml nitrobenzene concentration


Extraction temperature: 40 ºC; Extraction time: 30 min; Mixing speed: 600 rpm; Salt concentration: 30%, w:v


Desorption time: 3 min; Desorption temperature 250 ºC


Columns: CP-SIL (Varian, USA) 11 m ×0.25 mm ×0.25 µm


Inlet: 250 ºC, no shunt injection;


Carrier gas: nitrogen, flow rate 3 mL/min


Programmed temperature rise: 40 ºC for 2 min, 15 ºC/min to 160 ºC, 20 ºC/min to 250 ºC for 3 min


Detector: TSD or ECD


 

图片2.png 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

GC-TSD chromatogram

 

 

Linear range, linear correlation coefficient, minimum detection limit and repeatability

Analytes

Linear range (mg L-1)

Correlation coefficient (r)

LOD (ng/L)

RSD (n=5, %)


TSD

TSD

TSD

ECD

TSD

ECD

2 mg/L

10 mg/L

NB

0.1-10

0.9991

0.9991

0.9990

16

0.3

5.3

2.4

2-NT

0.1-10

0.9976

0.9976

0.9970

14

0.59

7.3

5.3

4- NT

0.1-10

0.9974

0.9974

0.9935

8

0.24

6.8

3.0

1,3-DNB

0.1-10

0.9982

0.9982

0.9980

38

0.43

5.7

1.1

2,6-DNT

0.1-10

0.9985

0.9985

0.9992

27

0.05

5.1

0.7

2,4-DNT

0.1-10

0.9992

0.9992

0.9997

22

0.06

5.5

0.9

TNT

0.5-10

0.9981

0.9981

0.9981

126

0.09

9.3

2.2

 

Actual Water Sample Test Results (TSD)

Analytes

Real sample

Recovery (%, RSD %, n=3) 


Seawater

Groundwater

Seawater

Groundwater

NB

ND

ND

103.3 (7.1)

90.7

2-NT

ND

0.81

91.7 (11.9)

90.7

4- NT

ND

ND

96.7 (8.5)

94.0

1,3-DNB

ND

ND

102.3 (6.4)

104.7

2,6-DNT

ND

ND

91.2 (6.2)

102.6

2,4-DNT

ND

ND

94.1 (6.5)

100.2

TNT

ND

5.36

100.3 (4.7)

114.9


Key word:Solidphasemicroextraction

1551752942177375.png

Qingdao Zhenzheng Analytical Instrument Co., LTD

Tel: 15963292045 Email: qdzhenzheng@163.com

QQ: 2776447184 Add: Building 27, Shandong University International Industrial Park, No.669 East China Road, High-tech Zone, Qingdao

Technical support:浩瀚网络     Record number:鲁ICP备14019093号-1  

Web address: www.qdzhenzheng.com           

热推产品  |  主营区域: 青岛 山东 广州 江苏 北京 浙江 湖北 重庆 上海

黄曲霉毒素检测

Please leave a message for us
Please input the message here, and we will contact you as soon as possible.
Full name
Contacts
Telephone
Seat / mobile phone number
mailbox
mailbox
address
address